SDKP Workstation for Pipettes Calibration
Calibration of automatic pipettes:
- with fixed volume
- with adjustable volume.
- table for calibration
- anti-vibration table
- probe for temperature, humidity and pressure measurement
- probe for water temperature measurement
- sink for used water
- armrests
- LCD monitor
- PC with pre-installed software PIPETTES RADWAG
- power supply
- keyboard and mouse
- container for used water
Pipette calibration adapters are intended for calibration and verification of piston pipettes using gravimetric method. Use of the pipette calibration adapter reduces maximum capacity of the balance by the mass of the weighing vessel. The table presents the minimum reguirements in accordance with ISO 8655 standard. It also enables to select suitable adapter for pipette calibration and respective balance version.
Selected volume1 of apparatus under test V |
Resolution mg |
Repeatability and linearity mg |
Recommended balance | ||||||||||
1 µl ≤ V ≤ 10 µl | 0,001 | 0,002 |
MYA 21.5Y PLUS
MYA 21.5Y
|
XA 6/21.5Y.M.A PLUS
XA 6/21.5Y.M.A
XA 6/21.5Y.M
|
|||||||||
10 µl ≤ V ≤ 100 µl | 0,01 | 0,02 |
XA 82/220.5Y.A PLUS
XA 82/220.5Y.A
XA 82/220.5Y
|
XA 120/250.5Y.A PLUS
XA 120/250.5Y.A
XA 120/250.5Y
|
|||||||||
100 µl ≤ V ≤ 1000 µl | 0,1 | 0,2 | |||||||||||
1 ml < V ≤ 10 ml | 0,1 | 0,2 | |||||||||||
Adapter for pipette calibration | MY 11 | XA 11 | XA 17 | XA 100 | XA 17 | XA 100 | |||||||
Maximum vessel capacity | 11 ml | 11 ml | 17 ml | 100 ml | 17 ml | 100 ml | |||||||
Weighing vessel mass | ~ 10 g | ~ 10 g | ~ 11 g | ~ 65 g | ~ 11 g | ~ 65 g |
If the standard uncertainty of measurement of the balance is known (e.g. from the balance calibration certificate), it may be used instead of the repeatability and linearity. The standard uncertainty of measurement shall not be more than two to three times the resolution [Reference: ISO8655-6:2002, pg. 2].