English

Español

AGV-8/1000 Automatic Comparator for determination of mass standard's density and volume

AGV-8/1000 Automatic Comparator for determination of mass standard's density and volume in Mass Comparators
This image is for reference only.

Product code: WL-418-0002
Request for quotation

Overview

Maximum capacity [Max]: 1110g
Readability [d]: 0.01mg

The AGV-8/1000 automatic comparator determines density and volume of

mass

standards of class E1 and lower in accordance with OIML R111. It is characterized by the highest accuracy and measurement stability ranging between 1g and 1kg.

The AGV-8/1000 automatic comparator enables density determination to be carried out simultaneously for 7

mass

standards using one reference weight.

The comparator features specially designed feeder to load and unload the magazine with mass standards. The device is equipped with an additional circulator eliminating air bubbles and a top-class liquid temperature sensor of 0.001 ˚C accuracy. A

weighing

pan suspended on wires eliminates the eccentricity and minimizes the influence of liquid surface tension.

Dedicated software enables to determine mass standards density and calculate uncertainty using A method according to OIML R111. It also allows to determine (verify) liquid density.

Videos

RMCS - Radwag Mass Comparator System

Datasheet

E1

1 g ÷ 1 kg

 

Maximum capacity [Max]

1110

 g

Readability [d]

0.01

 mg

Standard repeatability [5% Max]

0.04

 mg

Standard repeatability [Max]

0.05

 mg

Permissible repeatability

0.08

 mg

Electric compensation range

-10 g ÷ +110 g

 

Stabilization time

30

 s

Adjustment

external

 

Supplementary weights external

500 g, 300 g, 100 g, 100 g

 

Magazine positions

8

 

Power supply

100 ÷ 240 V AC 50 / 60 Hz

 

Operating temperature

+15 ÷ +30

 °C

Storage temperature

-20 ÷ +50

 °C

Operating temperature change rate

±0.5°C/12h (±0.3°C/4h)

 

Relative humidity

40% ÷ 60%

 

Relative humidity change rate

±5%/12h (3%/4h)

 

Weighing pan dimensions

ø60

 mm

Object diameters

5-94

 mm

Weighing device dimensions

690×710×1060

 mm

Controling device dimensions

206×140×70

 mm

Repeatability is expressed as a standard deviation determined for 6 ABBA cycles. Standard deviation is experimentally determined under ambient conditions for calibration of E1 class mass standards specified in OIML R111 (Table C.1.) document.

Documentation

Technical specification
Publications and teaching materials
Data Sheet
(0.90 MB)
Save See the publication
Automatic Comparison of Weights and Mass Standards 2017
(8.34 MB)
Save See the publication
Detection of vibrations at the electronic balances workstation
(2.02 MB)
Save See the publication
Mass Comparators
(11.13 MB)
Save See the publication
AGV Automatic Comparator
(3.06 MB)
Save See the publication

Accessories

RMCS System
RMCS Lite
Receipt Printer
THBR 2.0 System - Ambient Conditions Monitoring
RS 232 cables (scale - EPSON printer)
RS 232, RS 485 cables
Barcode scanners
☎ Call us now
✉ Get a quote

© 2018 RADWAG USA L.L.C.. All rights reserved.

All proprietor's rights do the materials placed in the server are in possession of RADWAG USA L.L.C. and are not to be copied, processed and utilized in commercial purpose unless agreed with RADWAG USA L.L.C. .

RADWAG USA L.L.C.
19599 NE 10th Ave., #E, North Miami Beach, FL, 33179 USA
tel.: 13056513522