| Maximum capacity [Max] |
25 g (per measuring channel) |
| Readability [d] |
0,01 mg |
| Standard repeatability [Max] |
10 µg |
| Standard repeatability [5% Max] |
5 µg |
| Linearity |
±0,03 mg |
| Stabilization time |
4 s |
| Adjustment |
internal (automatic), external |
Typical measurement time for a 12-channel pipette with constant volume, according to ISO 8655 (for d=10 µg)* |
5 min |
Typical measurement time for a 12-channel pipette with variable volume, according to ISO 8655 (for d=10µg)* |
15 min |
| Maximum volume |
16 ml (per measuring channel) |
| Display |
10″ graphic colour touchscreen |
| Delivery components |
pipette calibration balance, terminal, transport case, liquid suction pump, vapor curtain |
| Weighing pan dimensions |
intended |
| Device dimensions W x D x H |
563x393x214 |
| Packaging dimensions W x D x H |
490×700×700 mm |
| Net weight |
30 kg |
| Gross weight |
40 kg |
| Communication interface |
2×USB-A, USB-C, RS 232 (COM3), HDMI, Ethernet, Wi-Fi®, Hotspot |
| Power supply |
Adapter: 100 – 240V AC 50/60Hz 1A Max; 15V DC 2,4A Device: 12 – 15V DC 1,6A max; 10 – 29W* |
| Operating temperature |
+10 – +40 °C |
| Operating temperature change rate |
±0,3 °C / 1 h (±1 °C / 8 h) |
| Relative humidity |
20% – 80% |
| Relative humidity change rate |
±1% / h (±4% / 8 h) |