E1 Calibration Range |
- |
E2 Calibration Range |
- |
F1 Calibration Range |
- |
F2 Calibration Range |
- |
M1 Calibration Range |
1000 kg |
M2 Calibration Range |
200 – 1000 kg |
Maximum capacity [Max] |
1050 kg |
Readability [d] |
10 g |
Standard repeatability [5% Max] |
6 g |
Standard repeatability [Max] |
10 g |
Permissible repeatability |
15 g |
Eccentricity (tested load) |
1d / 5 mm |
Electric compensation range |
0 – 1050 kg |
Stabilization time |
10 s |
Adjustment |
internal (automatic) |
Display |
10″ graphic colour touchscreen |
Weighing pan dimensions |
1000×800 mm |
Controling device dimensions |
249×170×72 mm |
Comparator packaging dimensions without self-centering weighing pan |
1200×1000×440 mm |
Net weight |
121 kg |
Gross weight |
161 kg |
Recommended load speed |
0.3 m/min |
Maximum permissible load speed |
0.6 m/min |
Weighing unit dimensions without self-centering weighing pan |
1010×800×182 mm |
Communication interface |
2×USB-A, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot |
Operating temperature |
+15 – +30 °C |
Operating temperature change rate |
±2 °C / 12 h |
Relative humidity |
40% – 60% |
Relative humidity change rate |
±10% / 4 h |
is expressed as a standard deviation determined for 6 ABBA cycles. Standard deviation is experimentally determined under ambient conditions for calibration of F2 class mass standards specified in OIML R111 (Table C.1.) document.